Fluke 724 Temperature Calibrator
Temperature Calibrators
An easy-to-use temperature calibrator that can measure and source 12 thermocouple and seven RTD types, volts, and ohms, which allows users to test and calibrate a wide variety of temperature instruments. Additionally, this tool also handles high-speed pulsed RTD circuits and offers loop power.
| Stock No. | FLK5131125 |
|---|---|
| Manufacturer | Fluke |
| Part No. | 724 |
| Supplied/Stocked By | FLUKE |
Request a quote or more info
Fluke 724 Offers
Test and calibrate all temperature sensors and transmitters in your plant with only this temperature calibrator. This instrument is easy-to-use, and sources and measures 12 thermocouple and seven RTD types, plus volts and ohms.
Additionally, this tool as well handles high-speed pulsed RTD circuits and provides loop power. It features a dual-line measure/source display that allows users to source temperature and view loop current simultaneously and a "no menus" controls.
Features
- Easy to read dual display lets you view input and output simultaneously
- Measure RTDs, thermocouples, ohms, and volts to test sensors and transmitters
- Source/simulate thermocouples, RTDs, volts, and ohms to calibrate transmitters
- Perform fast linearity tests with 25% and 100% steps
- Execute remote tests with auto step and auto ramp
- Power transmitters during test using loop power supply with simultaneous mA measurement
- Store frequently-used test setups for later use
- Backlight lets you work in poor light
- Large battery capacity of four AA cells
- Battery door for easy changes
- A compact, streamlined shape that makes it easy to carry
- A rugged, reliable design that stands up to field use
What's included with the Fluke 724
- Calibrator
- NIST Traceable Calibration Report and Data
- Test Clips
- Test Leads
- Users Manual
- Users Manuals on CD-ROM
| Measurement Accuracy | |
|---|---|
| Voltage DC | 30.000 V: 0.02% + 2 counts (upper display)20.000 V: 0.02% + 2 counts (lower display)100.00 mV: 0.02% + 2 counts-10.00 to 75.00 mV: 0.025% + 1 count (via TC connector) |
| Current DC | 24.000 mA: 0.02% + 2 counts |
| Resistance | 0 to 400.0 Ω: 0.1 Ω (4-wire), 0.15 Ω (2- and 3-wire)401 to 1500 Ω: 0.5 Ω (4-wire), 1 Ω (2- and 3-wire)1500 to 3200 Ω: 1 Ω (4-wire), 1.5 Ω (2- and 3-wire) |
| Source Accuracy | |
| Voltage DC | 100.00 mV: 0.02% +2 counts10.000 V: 0.02% +2 counts–10.00 to 75.00 mV: 0.025% + 1 count (via TC connector) |
| Resistance | 15.0 to 400.0 Ω: 0.15 Ω (exc. current 0.15 to 0.5 mA), 0.1 Ω (exc. current 0.5 to 2 mA)401 to 1500 Ω: 0.5 Ω (excitation current 0.05 to 0.8 mA)1500 to 3200 Ω: 1 Ω (excit... |
| Functions | |
| Ramp Functions | Source functions: Voltage, current, resistance, frequency, temperatureRamps: Slow ramp, Fast ramp, 25% step-ramp |
| Loop Power Function | Voltage: 24 VAccuracy: 10%Maximum current: 22 mA, short circuit protected |
| Step Functions | Source functions: voltage, resistance, temperatureSteps: 25% of range, 100% of range |
| Environmental Specifications | |
| Operating Temperature | 14 to 131°F (–10 to 55°C) |
| Storage Temperature | -4 to 160°F (–20 to 71°C ) |
| Humidity (Without Condensation) | 90% = 50 to 86°F (10 to 30°C)75% = 86 to 104°F (30 to 40°C)45% = 104 to 122°F (40 to 50°C)35% = 122 to 131°F (50 to 55°C) |
| Safety Specifications | |
| Safety Rating | CSA C22.2 No. 1010.1:1992 |
| EMC | EN50082-1:1992 and EN55022:1994 Class B |