Fluke 7211603 Dual Sensor Pressure Calibrator
Pressure Calibrators
EGP 332,955.00/Each
(ex VAT)
EGP 382,898.25/Each
(inc VAT)
Precision Pressure Calibrator with dual isolated sensors for gas custody transfer applications. Ranges: 16 psi (low) 300 psi (high).
Stock No. | FLK4353283 |
---|---|
Manufacturer | Fluke |
Part No. | 721-1603 |
Supplied/Stocked By | FLUKE |

Fluke 721-1603 Offers
Features
- Two isolated, stainless steel, pressure sensors with 0.025% f.s. accuracy
- Can be configured by selecting either a 16 psig (1.1 bar) or 36 psig (2.48 bar) low pressure sensor and then adding your choice of 7 high pressure ranges
- External pressure module connection supports all pressure modules (29 ranges) for even more measurement capability
- Temperature-compensation ensures accuracy in field applications
- Pt100 RTD input for temperature measurement, accurate to 0.1°C (0.2°F) (Requires optional RTD probe sold separately)
- Measure 4-20 mA signals
- Internal 24 V Loop Power Supply can power a transmitter under test
- Measure up to 30 V DC
- Large graphic style LCD with backlight can be configured to display up to 3 inputs simultaneously (i.e. pressure 1, pressure 2, and RTD or any combination of inputs)
- Pressure displayed in the user's choice of 19 engineering units
- Switch test on both pressure inputs, or external pressure module
- Operating mode layers support simple measurements to complex applications
- Up to five frequently used setups can be stored; last setup automatically recalled on power-up
- % error and damping functions
Application Selection Guide
Pressure Calibration Selection GuideWhat's Included with the Fluke 721 Series
- Dual Sensor Pressure Calibrator
- Soft case
- Four (4) AA batteries
- Manual
- Traceable calibration certificate
- Test leads
Specifications 1-Year | |
---|---|
Low pressure sensor | Range |
-14 psi to +16 psi, -0.97 bar to 1.1 bar | 0.001 psi, 0.0001 bar |
High pressure sensor | -12 psi to +300 psi, -0.83 bar to 20 bar |
General Specifications | |
Instrument setup recall | Last used on power-up |
Environmental | Operating temperature -10 °C to +50 °C Storage temperature (without batteries) -20 °C to +60 °C |
Altitude | 2,000 meters |
Ingress protection | IP-40 |
Power requirements | 6.0 v dc |
Battery life | >35 hours, typical usage |
Batteries required | (4) Standard AA cells |
Dimentions (HxWxD) | 7.9" X 4.3" X 2.3" (20 cm X 11 cm X 5.8 cm) |
Weight | 03539 kg (1 lb 3 oz) |
Connectors/ports pressure | Two, 1/8 in NPT, 700P Series Pressure Module; RTD Probe |
Electrical and Temperature Measurement 1-Year | |
Function | Range |
mA measure | 0 to 24 mA |
VDC measure | 0 to 30 VDC |
*Temperature measure (RTD/ Ohms) | -40 °C to 150 °C (-40 °F to 302 °F) |
Loop power supply | 24V |
*Temperature measurement requires optional Fluke-720RTD Pt-100 RTD probe available as an accessor... | |
Temperature Effect | |
No effect on accuracy on all functions from 15 °C to 35 °C | |
Add ± 0.002 % F.S./°C for temps outside of 15 °C to 35 °C | |
Available Pressure Engineering Units | |
psi, bar, mbar, kPa, kgcm2, cmH2O@4°C, cmH2O@20°C, mH2O@4°C, mH2O@20°C, inH2O@4&d... |
Parts List